Measuring Instruments

Multi-angle spectrophotometer for metallic colors.
| Illuminating/viewing system | 3-angle circumferential illumination (25°, 45°, and 75°)/0° viewing angle |
|---|---|
| Wavelength range | 400 to 700nm |
| Wavelength pitch | 20nm |
| Photometric range | 25°: 0 to 300% reflectance; 45°, 75°: 0 to 200% reflectance; Resolution: 0.01% |
| Light source | Pulsed xenon arc lamp |
| Minimum measurement interval | 7 seconds (when measuring a white calibration plate at 23°C) |
| Battery life | Alkaline manganese batteries: Approx. 400 measurements at 10 sec. intervals (when dark color is measured) |
| Measurement/illumination area | 12mm/ 20mm |
| Repeatability | Spectral reflectance: Standard deviation within 0.30% Colorimetric values: Standard deviation ΔE*ab within 0.05 (Measurement conditions: White calibration plate measured 30 times at 10s intervals) |
| Interface | RS-232C Terminal; D-sub 9-pin (female) |
| Display | Dot-matrix reflective LCD with 21 characters ×7 lines (128 × 56 dots) With contrast control slide |
| Display data | Colorimetric data: L*a*b*, L*C*h Color difference data: Δ(L*a*b*), Δ(L*C*H*), ΔE*ab, CMC(l:c), ΔE00(CIE DE2000) Other data display: FF value, line graph, temperature |
| Illuminant/Observer conditions | Light source: A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12 Observer: CIE 2° and 10° |
| Temperature detection | Detector: Thermopile Wavelength: 8 to 13µm Measurement/display range: -10 to 80°C (0.1°C increments) Measurement diameter: Φ20mm Measurement rating: 0 to 50°C: ±2.5°C ±1 digit (when an object of 0.93 emissivity is measured at ambient temperatures of 18 to 28°C) |
| Operating temperature/humidity range | 0 to 40°C; relative humidity 85% or less (at 35°C) with no condensation |
| Storage temperature/humidity range | -20 to 45°C; relative humidity 85% or less (at 35°C) with no condensation |
| Power | Four AA-size batteries or special AC adapter |
| Dimensions (W × H × D) | 115 × 257 × 164mm |
| Weight | Approx.1.4 kg (without batteries) |
Specifications and drawings are subject to change without prior notice.