The essentials of imaging

  • Global
  • Measuring Instruments
search form

Color Analyzers

Measuring Instruments

content

CA-210

Display Color Analyzer

Faster, more sensitive to low brightness, more accurate, and easier to use.

Specifications

Principal Specifications

Receptor Silicon photocell
Measurement area Φ27mm
Acceptance angle ±2.5°; Pointing function: By LED
Measurement distance 30±10mm
Display range Luminance 0.01 to 1000cd/m2
Chromaticity Displayed in 4- or 3-digit value (selectable)
Luminance Measurement range 0.10 to 1000cd/m2
Accuracy (for white) *1 ±2%, ±1 digit of reading (Temperature: 23°C ±2°C; Relative humidity:(40±10)%)
Repeatability (2σ) *1 0.10 to 0.99cd/m2: 0.2%+1digit; 1.0 to 1000cd/m2: 0.1%+1digit
Chromaticity Measurement range 0.10 to 1000cd/m2
Accuracy *1 0.10 to 4.99cd/m2: ±0.008 for white
5.00 to 39.99cd/m2: ±0.005 for white
40.00 to 1000 cd/m2: ±0.003 for white
160 cd/m2: ±0.002 for white; ± 0.004 for monochrome *2
Repeatability (2σ) *1 0.10 to 0.19cd/m2: 0.015
0.20 to 0.49cd/m2: 0.008
0.50 to 1.99cd/m2: 0.003
2.00 to 1000 cd/m2: 0.001
Measurement speed (meas./sec.) xyLv*3 0.10 to 1.99 cd/m2: 5 (4.5)
4 cd/m2 or greater: 20 (17)
Display Digital xyLv, XYZ, TΔuvLv, u'v'Lv, RGB analyze
Analog ΔxΔyΔLv, R/G B/G ΔG, ΔR B/R G/R
LCD 16 characters × 2 lines (with backlight)
SYNC mode NTSC, PAL, EXT, UNIV, INT
Measurement subject Vertical synchronous frequency: 40 to 200Hz
Memory channels 100
Analyzer function Standard function
Interface USB (Rev. 1.1) or RS-232c (38,400 bps maximum)
Multi-point measurement Max. 5 points (using optional 4-Probe Expansion Board CA-B14)
Software Solutions SDK software (supplied as standard accessory)
Operating temperature/humidity range Temperature: 10 to 28°C; Relative humidity 70% or less with no condensation. Luminance change: ±2% ±1 digit of reading for white; Chromaticity change: ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's standard LCD *1, 160.0 cd/m2 at 23°C 40%RH
Storage temperature/humidity range 0 to 28°C: Relative humidity 70% or less with no condensation; 28 to 40°C: Relative humidity 40% or less with no condensation
Input voltage range 100 to 240V~50/60Hz, 50VA
Dimensions (W × H × D) Main body: 340 × 127 × 216mm
Probe: Φ49 × 204mm
Weight Main body: 3.58kg; Probe: 520g

For more details, or for details regarding Small Universal Measuring Probe and LCD Flicker Measuring Probes, please request a catalog from your sales represenatative.

*1:
Chromaticity and luminance measured under Konica Minolta's conditions (Standard LCD at 6500K and 9300K)
*2:
The luminance for monochrome is measured when the reading of luminance for white is 160 cd/m2.
*3:
For single probe and USB; figures in parentheses for RS-232C with baud rate of 38,400 bps.
*
Specifications are subject to change without notice.

Back to Top

©2007-2009 Konica Minolta Sensing, Inc.