Spectroradiometer CS-2000

NEW

Highly precise spectral radiance/chromaticity measurement possible from 0.003 cd/m2
Grand Prize Award, Advanced Display of the Year 2008 prize

Specifications

Major specifications of the CS-2000

Model CS-2000
Wavelength 380 to 780 nm
Wavelength 0.9 nm/pixel
Display wavelength bandwidth 1.0 nm
Wavelength precision ±0.3 nm (Median wavelength: 435.8 nm, 546.1 nm, 643.8 nm; Hg-Cd lamp)
Spectral bandwidth 5 nm or less (half bandwidth)
Measuring angle (selectable) 0.2° 0.1°
Measurement luminance range (Standard light source A) 0.003 to 5,000 cd/m2 0.075 to 125,000 cd/m2 0.3 to 500,000 cd/m2
Minimum measuring area φ5 mm (φ1 mm when using close-up lens) φ1 mm (φ0.2 mm when using close-up lens) φ0.5 mm (φ0.1 mm when using close-up lens)
Minimum measuring distance 350 mm (55 mm when using close-up lens)
Minimum luminance display 0.00002 cd/m2
Accuracy: Luminance *1(Standard light source A) ±2%
Accuracy: Chromaticity *1(Standard light source A) x,y :±0.003 (0.003 to 0.005 cd/m2)
x,y :±0.002 (0.005 to 0.05 cd/m2)
x :±0.0015 (0.05 cd/m2 to )
y :±0.001
x,y :±0.003 (0.075 to 0.125 cd/m2)
x,y :±0.002 (0.125 to 1.25 cd/m2)
x :±0.0015 (1.25 cd/m2 to )
y :±0.001
x,y :±0.003 (0.3 to 0.5 cd/m2)
x,y :±0.002 (0.5 to 5 cd/m2)
x :±0.0015 (5 cd/m2 to )
y :±0.001
Repeatability: Luminance(2σ) *2 0.4 % (0.003 to 0.05 cd/m2)
0.3 % (0.05 to 0.1 cd/m2)
0.15 % (0.1 to 5,000 cd/m2)
0.4 % (0.075 to 1.25 cd/m2)
0.3 % (1.25 to 2.5 cd/m2)
0.15 % (2.5 to 125,000 cd/m2)
0.4 % (0.3 to 5cd/m2)
0.3 % (5 to 10cd/m2)
0.15 % (10 to 500,000cd/m2)
Repeatability: Chromaticity(2σ) *2 0.002 (0.003 to 0.005 cd/m2)
0.001 (0.005 to 0.1 cd/m2)
0.0006 (0.1 to 0.2 cd/m2)
0.0004 (0.2 to 5,000 cd/m2)
0.002 (0.075 to 0.125 cd/m2)
0.001 (0.125 to 2.5 cd/m2)
0.0006 (2.5 to 5 cd/m2)
0.0004 (5 to 125,000 cd/m2)
0.002 (0.3 to 0.5 cd/m2)
0.001 (0.5 to 10 cd/m2)
0.0006 (10 to 20 cd/m2)
0.0004 (20 to 500,000 cd/m2)
Polarization error 2 % or less(400 to 780 nm):1°, 3 % or less(400 to 780 nm):0.1° 0.2°
Integration time Fast: 0.005 to 16 sec.; Normal: 0.005 to 120 sec.
Measurement time 1 sec. minimum (Manual mode) to 243 sec. maximum (Normal mode)
Color space mode Lv x y, Lv u' v', Lv TΔuv, XYZ, spectral graph, dominant wavelength, excitation purity
Interface USB 1.1
Operating temperature/
humidity range
5 to 35 °C, relative humidity 80% or less with no condensation
Storage temperature
/humidity range
0 to 35 °C, relative humidity 80% or less with no condensation
Power AC adapter (100 to 240V~,50/60 Hz)
Current consumption Approx. 20 W
Size 158 (W) x 262 (H) x 392 (D) mm
Weight 6.2 kg
*1
Average of 10 measurements in Normal mode at a temperature of 23±2°C and a relative humidity of 65% or less.
*2
10 measurements in Normal mode at a temperature of 23±2°C and a relative humidity of 65% or less.

System Configuration

System Configuration

Data Management Software

CS-S10w Professional (Standard accessory)

With this software, the CS-2000 can be controlled from a personal computer to display measured data in various graphs or lists, to transfer data to spreadsheet software, or to copy-and-paste data. CS-S10w offers various data management, analysis and evaluation options to assist in research and development or quality control.


  • The specifications and drawings given here are subject to change without prior notice.