X-Ray Generator Specifications:
|
Power Rating (kW): DIN
|
80 kW / 65 kW
|
50 kW
|
|
Ultra High Frequency:
|
120 kHz PLUS
|
120 kHz PLUS
|
|
kVp Range (1 kVp Steps):
|
40 - 150 kVp
|
40 - 125 kVp
|
|
150 kVp Output:
|
Standard
|
Optional
|
|
mAs Range:
|
0.025 - 800 mAs
|
0.025 - 800 mAs
|
|
mA Range:
|
25 - 800 mA
|
25 - 690 mA
|
|
Timer Range:
|
0.001 - 6.3 (seconds)
|
0.001 - 6.3 (seconds)
|
|
High (Dual) Speed Starter:
|
Standard
|
Standard
|
|
Console:
|
Anatomical Programming (APR 110 views / 5000 techniques)
|
Anatomical Programming (APR 110 views / 5000 techniques)
|
Automatic Exposure Control: (3-field AEC)
|
Standard
|
Standard
|
|
Power (Three Phase Line):
|
380 - 480 VAC +/- 10%
|
380 - 480 VAC +/- 10% standard 280 VAC +/- 5% optional
|
RDIS Elevating Float Top Radiographic Table
- Patient weight capacity: 650 lbs. (295.5 kg)
- Elevating range: 21" – 32.5" (53 – 83 cm)
- Collision avoidance electronics with built-in table sensors
- Tabletop length: 85" (216 cm); with 32" (82 cm) of longitudinal travel
- Tabletop width: 35.5" (90 cm); with 10" (25 cm) of transverse travel
- Recessed foot switches for all table movements, with safety lock-out control switch
- Float-top multi-function hand control switch, adjustable position (for elevation and float control)
- Adjustable patient handgrips along concealed accessory rails
Options:
- Compression Band
- Lateral Cassette Holder
RDIS Floor Mounted Tubestand
- 10 ft. long tracks with 98" (249 cm) of longitudinal travel
- ALL-LOCKS release switch and auto-stop sensors for horizontal and vertical positioning
- Vertical travel: 60.5" (154 cm), minimum floor to focus distance of 13.75" (35 cm)
- Column rotation: +/-180º, Transverse Arm: 10" (25 cm)
- Tube angulations: +/-130º with detents at 0º and +/-90º
- Available with Trunnion mount (QS-55T) which allows for tube head rotation +45º/-20º with dual angle guide
RDIS Ceiling Mounted Tube Support
- Telescopic column with vertical travel: 59" (150 cm)
- Longitudinal travel: 140" (356 cm) with rail length of 14' (4.25 meters)
- Transverse travel: 87.5" (222 cm) with rail length of 10' (3 meters)
- Tube rotation about vertical axis: +154º/-182º, detents at 0º and +/-90º
- Tube rotation about horizontal axis: +/-120º, detents at 0º and +/-90º
- Operator handgrips with digital display, dual “All Lock” release switch in grips