The essentials of imaging

search form

REGIUS Digital Imaging Suite

Medical Imaging

content

REGIUS Digital Imaging Suite

REGIUS Digital Imaging Suite

X-Ray Generator Specifications:

Power Rating (kW): DIN 80 kW / 65 kW 50 kW
Ultra High Frequency: 120 kHz PLUS 120 kHz PLUS
kVp Range (1 kVp Steps): 40 - 150 kVp 40 - 125 kVp
150 kVp Output: Standard Optional
mAs Range: 0.025 - 800 mAs 0.025 - 800 mAs
mA Range: 25 - 800 mA 25 - 690 mA
Timer Range: 0.001 - 6.3 (seconds) 0.001 - 6.3 (seconds)
High (Dual) Speed Starter: Standard Standard
Console: Anatomical Programming
(APR 110 views / 5000 techniques)
Anatomical Programming
(APR 110 views / 5000 techniques)
Automatic Exposure Control:
(3-field AEC)
Standard Standard
Power (Three Phase Line): 380 - 480 VAC +/- 10% 380 - 480 VAC +/- 10% standard
280 VAC +/- 5% optional

RDIS Elevating Float Top Radiographic Table

  • Patient weight capacity: 650 lbs. (295.5 kg)
  • Elevating range: 21" – 32.5" (53 – 83 cm)
  • Collision avoidance electronics with built-in table sensors
  • Tabletop length: 85" (216 cm); with 32" (82 cm) of longitudinal travel
  • Tabletop width: 35.5" (90 cm); with 10" (25 cm) of transverse travel
  • Recessed foot switches for all table movements, with safety lock-out control switch
  • Float-top multi-function hand control switch, adjustable position (for elevation and float control)
  • Adjustable patient handgrips along concealed accessory rails

Options:

  • Compression Band
  • Lateral Cassette Holder

RDIS Floor Mounted Tubestand

  • 10 ft. long tracks with 98" (249 cm) of longitudinal travel
  • ALL-LOCKS release switch and auto-stop sensors for horizontal and vertical positioning
  • Vertical travel: 60.5" (154 cm), minimum floor to focus distance of 13.75" (35 cm)
  • Column rotation: +/-180º, Transverse Arm: 10" (25 cm)
  • Tube angulations: +/-130º with detents at 0º and +/-90º
  • Available with Trunnion mount (QS-55T) which allows for tube head rotation +45º/-20º with dual angle guide

RDIS Ceiling Mounted Tube Support

  • Telescopic column with vertical travel: 59" (150 cm)
  • Longitudinal travel: 140" (356 cm) with rail length of 14' (4.25 meters)
  • Transverse travel: 87.5" (222 cm) with rail length of 10' (3 meters)
  • Tube rotation about vertical axis: +154º/-182º, detents at 0º and +/-90º
  • Tube rotation about horizontal axis: +/-120º, detents at 0º and +/-90º
  • Operator handgrips with digital display, dual “All Lock” release switch in grips

Back to Top

©2007-2011 Konica Minolta Medical Imaging U.S.A., Inc.