Measuring Instruments

KONICA MINOLTA RANGE7, a high-accuracy instrument for measuring the three-dimensional shape of subjects such as forged or cast parts, press-formed parts, plastic parts, etc.
| Measuring method | Triangulation light block method | ||
|---|---|---|---|
| Light source | Semiconductor laser; Wavelength: 660nm | ||
| Laser class | Class 2 (IEC60825-1 Amendment 2) | ||
| Number of pixels taken | 1.31 megapixels (1280×1024) | ||
| Measurement distance | 450mm to 800mm | ||
| Receiving lens | TELE; WIDE | ||
| Measurement range (Units: mm) | X × Y | Z | |
| TELE; Distance: 450mm | 79×99 | 54 | |
| TELE; Distance: 800mm | 141×176 | 97 | |
| WIDE; Distance: 450mm | 150×188 | 109 | |
| WIDE; Distance: 800mm | 267×334 | 194 | |
| XY direction measurement interval (Units: mm) | TELE; Distance: 450mm | 0.08 | |
| TELE; Distance: 800mm | 0.14 | ||
| WIDE; Distance: 450mm | 0.16 | ||
| WIDE; Distance: 800mm | 0.28 | ||
| Accuracy (Distance between balls)*1 | ±40µm | ||
| Precision (Z, s)*2 | 4µm | ||
| Autofocus | Yes | ||
| Autoexposure | Yes | ||
| Scan time | Approx. 2 sec. (1 scan) | ||
| Preview function | Yes (Approx. 0.4 sec./scan) | ||
| Ambient lighting condition | 500lx or less | ||
| File format*3 | Input/output: Konica Minolta proprietary formats .rgv, .rvm; Output only: .stl | ||
| Output interface | USB 2.0 High Speed | ||
| Power | Included AC Adapter | ||
| Input voltage: AC 100 to 240V (50/60Hz) | |||
| Rating: 1.4A (at AC 100V input) | |||
| Dimensions | 295 (W) × 190 (H) × 200 (D) mm (Not including grips and lens) | ||
| Weight | Approx. 6.7kg | ||
| Operating temperature/humidity range | 10 to 40°C; Relative humidity 65% or less with no condensation | ||
| Storage temperature range | -10 to 40°C; Relative humidity 85% (at 35°C) or less with no condensation | ||
| *1 | When measuring distance between balls for a ball bar as defined in VDI/VDE 2634-2 under Konica Minolta measurement conditions. Konica Minolta measurement conditions: Temperature: 20±1°C; Lens: TELE; Measurement distance: 450mm; Warmup: 20 min.; Software used: Konica Minolta processing software; After calibration of instrument; Measurement subject: Konica Minolta standard (2 balls); Arrangement of measurement subject: Konica Minolta standard arrangement (10 locations inside measurement space); Uncertainty of standard not included |
|---|---|
| *2 | Measurement conditions: Temperature: 20±1°C; Lens: TELE; Measurement distance: 450mm; Warmup: 20 min.; Measurement subject: Konica Minolta standard planar chart; Software used: Konica Minolta processing software; 1σ value |
| *3 | Using Konica Minolta processing software |
| Readable formats | Konica Minolta proprietary formats .rgv (Single-scan data cloud), .rvm (multiple data) |
|---|---|
| Data output | ASCII point group; Binary point group (including normal vectors) |
| Konica Minolta proprietary formats .rgv, .rvm | |
| Measurement functions | Monitor image; Preview; AF/AE; Measurement |
| Control of rotary stage (Konica Minolta optional accessory) | |
| Editing functions | Data alignment; Data merging; Point group deletion |
| Drawing | Point group shading |
| OS | Windows Vista 64-bit (x64) |
|---|---|
| CPU | Core2Duo, Xeon, or better |
| RAM | 4GB or more |
| Display | Graphics display capability of 1280×1024 or more |
| Graphics board | OpenGL compatible board (verified-compatible board recommended) |
| Interface | USB 2.0 port |