CM-2600d / 2500d

CM-2600d / 2500d

Spectrophotometer (spectral type)

A portable device providing the performance and convenience of a benchtop spectrophotometer in a compact package.

Specifications

Principal Specifications

Model CM-2600d CM-2500d
Illumination / obser vation system d/8 (diffuse illumination, 8-degree viewing ), equipped with simultaneous measurement of SCI (specular component included)/SCE (specular component excluded) Conforms to CIE No.15,ISO 7724/1,ASTM E1164,DIN 5033 Teil7 and JIS Z8722 Condition C standard.
Sphere Size Φ52mm
Light-receiving element Silicon photodiode array (dual 40 elements)
Spectral separation device Diffraction grating
Wavelength range 360nm to 740nm
Wavelength pitch 10nm
Half bandwidth Approx. 10nm
Reflectance range 0 to 175%, resolution: 0.01%
Light source 3 pulsed xenon lamps 2 pulsed xenon lamps
Measurement time Approx. 1.5 seconds (approx. 2 seconds for fluorescent measurement) Approx. 1.5 seconds
Minim um interval between measurements 3 seconds for SCI/SCE (4 seconds for fluorescent measurement) 3 seconds for SCI/SCE
Battery perfomance Alkaline manganese:approx. 1000 measurements
Measurement/illumination area

MAV : Φ8mm/Φ11mm

SAV : Φ3mm/6mm (Selectable between MAV and SAV)

MAV : Φ8mm/Φ11mm
Repeatability

Spectral Reflectance:Standard deviation within 0.1% (360 to 380nm within 0.2%)

Colorimetric Value : Standard deviation within ΔE*ab 0.04(Measurement conditions:White calibration plate measured 30 times at 10-second intervals after white calibration was performed)

Inter instrument agreement withinΔE*ab 0.2 (MAV/SCI) Average for 12BCRA Series II color tiles compared to values measured with master body.
UV adjustment Instantaneous numerical adjustment (no mechanical adjustment required) With UV400nm cut filter No UV adjustment
Measurement mode Single measurement/automatic averaging of multiple measurements(auto mode: 3, 5, 8 times/manual mode)
Interface RS-232C standard
Observer 2/10 degrees (CIE 1931/2°,CIE 1964/10°)
Illuminant A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12 (simultaneous evaluation is possible using two light sources)
Display data Spectral value/graph, colorimetric value, color difference value/graph, PASS/FAIL result
Color space / colorimetric data L*a*b*, L*C*h, CMC (1:1), CMC (2:1), CIE94, Hunter Lab, Yxy, Munsell, XYZ, MI, WI (ASTM E313), YI (ASTM E313/ASTM D1925), ISO Brightness (ISO 2470), Density statusA /T, WI/Tint (CIE/Ganz), CIE00
Data memory

1700 pieces of data (as SCI/SCE 1 data) * 700 pieces of data in the "defined in COND." mode.

* Total of the sample data for the COND and TASK modes and color diference target data

Tolerance judgment Tolerance for color difference (both box and eliptical tolerances can be set)
Power source 4 AA-size battery or AC adapter
Size (W×H×D) 69 × 96 × 193mm
Weight Approx. 670g (without batteries)
Operating temperature / humidity range (*1) 5 to 40°C, relative humidity 80% or less (at 35°C) with no condensation
Storage temperature/humidity range 0 to 45°C, relative humidity 80% or (at 35°C) with no condensation
Standard accessories White calibration plate, Target mask Φ8mm, Target mask Φ3mm , RS-232C cable, AC adapter,AA-size battery (×4)
Optional Accessories Hard case, Dust cover set, Dust cover, SpectraMagic™NX (software), Zero calibration box
*1 Operating temperature/humidity range of products for North America : 5 to 40°C, relative humidity 80% or less (at 31°C) with no condensation