The essentials of imaging

search form

Luminance & Color Meters

Measuring Instruments

content

CS-200

CS-200

Luminance & Color Meters

CS-200 ChromaMeter measures luminance and chromaticity of light emitting products as diverse as large plasma displays, compact LCDs and LEDs, outdoor screens, high pressure lamps, and instrument panels.

Specifications

Principal Specifications

Model CS-200
Measurement range 0.01 - 200,000 cd/m2 (Measuring angle 1°)
0.01 - 5,000,000 cd/m2 (Measuring angle 0.2°)
0.01 - 20,000,000 cd/m2 (Measuring angle 0.1°)
Accuracy
(Measuring angle 1°) #1
(Temperature: 23°C±2°C, Relative humidity: 65% max.)
150 cd/m2 ( for Illuminant A ) : Lv ±2 % ±1digit ; xy ±0.002
0.01-0.5 cd/m2 ( for Illuminant A ) : Lv ±0.02cd/m2 ±1digit ; ---
0.5-1 cd/m2 ( for Illuminant A ) : Lv ±0.02 cd/m2 ±1digit ; xy ±0.007
1-10 cd/m2 ( for Illuminant A ) : Lv ±2 % ±1digit ; xy ±0.004
10-200,000 cd/m2 ( for Illuminant A) : Lv ±2 % ±1digit ; xy ±0.003
5000 cd/m2 ( for Illuminant A ) : color filter (R, G, B) ; xy ±0.006
Repeatability
(Measuring angle 1°) #2
0.01-1 cd/m2 ( for Illuminant A ) : Lv ±0.01 cd/m2 ±1digit ; --- (2σ/AUTO)
1-2 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.002 (2σ/AUTO)
2-4 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.001 (2σ/AUTO)
4-8 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.0005 (2σ/AUTO)
8-200,000 cd/m2 ( for Illuminant A ) : Lv ±0.1% ±1digit ; xy 0.0004 (2σ/AUTO)
Measurement time Automatic setting between 1s and 60s (AUTO); Automatic setting to 1s or 3s (LTD. AUTO)
0.5 sec / meas. (Super-FAST) 1 sec / meas.(FAST)3 sec / meas. (SLOW) 12 sec / meas. (Super-SLOW)
Measurement method Spectral method, Grating + linear photo diode array
Measuring angle 1°, 0.2°, 0.1°(switchable)
Minimum measuring area 0.5 mm
0.1 mm (close up lens)
Minimum measuring distance 296 mm (Distance from front edge of metal lens barrel)
Observer 2/10 degrees
Color space Lv x y, Lv u' v', Lv TΔuv, XYZ, dominant wavelength
Measurement synchronization setting range Vertical synchronization frequency : 40.00 to 200.00Hz
Interface USB 1.1
Power source AC adapter or 4 AA-Size Batteries
Battery life Approx. 3 hours (continuous measurement / Fast mode / AA-size alkaline cells)
Size 95 mm (W) × 127 mm (H) × 330 mm (L)
Weight 1.8 kg (without battery)
Operating temperature/humidity range 0°C to 40°C, relative humidity 85% or less (at 35°C) with no condensation
Storage temperature /humidity range 0°Cto 45°C,relative humidity 85% or less (at 35°C)with no condensation
#1 23°C±2C Lv = 0.01-10 cd/m2, SLOW, average of 30 measurements
Lv = 10 cd/m2 and higher, SLOW, average of 10 measurements
#2 At 0.2°measuring angle, the amount of received light is approx. 1/25 of that for 1°.
Therefore, the repeatability becomes the same as that for 1°with 25 times lower luminance.
At 0.1°measuring angle, the amount of received light is approx. 1/100 of that for 1°.
Therefore, the repeatability becomes the same as that for 1°with 100 times lower luminance.
* Specifications are subject to change without notice.

Back to Top