LS-100/LS-110

LS-100/LS-110

Luminance Meters

A lightweight, compact, highly-portable luminance meter for measurements of a wide range of luminance conditions.

Specifications

Principal Specifications

Model Luminance Meter LS-100 Luminance Meter LS-110
Type SLR spot luminance meter for measuring light-source and surface brightness
Acceptance angle 1/3°
Optical system 85mm f/2.8 lens; SLR viewing system; flare factor less than 1.5%
Angle of view
Focusing distance 1014mm (40 in.) to infinity
Minimum measuring area Φ14.4mm Φ4.8mm
Receptor Silicon photocell
Relative Spectral Response* Within 8% (f1卒) of the CIE spectral luminous efficiency V (λ
Response time FAST: Sampling time: 0.1s, time to display: 0.8 to 1.0s; SLOW: Sampling time: 0.4s, time to display: 1.4 to 1.6s
Luminance units cd/m2 or fL (switchable)
Measuring range

FAST: 0.001 to 299,900cd/m2 (0.001 to 87,530fL)

SLOW: 0.001 to 49,990cd/m2 (0.001 to 14,590fL)

FAST: 0.01 to 999,900cd/m2 (0.01 to 291,800fL)

SLOW: 0.01 to 499,900cd/m2 (0.01 to 145,900fL)

Accuracy

0.001 to 0.999cd/m2 (or fL): ±2% ±2 digits of displayed value

1.000cd/m2 (or fL) or greater: ±2% ±1 digit of displayed value

0.01 to 9.99cd/m2 (or fL): ±2% ±2 digits of displayed value

10.00cd/m2 (or fL) or greater: ±2% ±1 digit of displayed value

(Illuminant A measured at ambient temperature of 20 to 30°C/68 to 86°F)
Repeatability

0.001 to 0.999cd/m2 (or fL): ±0.2% ±2 digits of displayed value

1.000cd/m2 (or fL) or greater: ±0.2% ±1 digit of displayed value

0.01 to 9.99cd/m2 (or fL): ±0.2% ±2 digits of displayed value

10.00cd/m2 (or fL) or greater: ±0.2% ±1 digit of displayed value

(Measurement subject: Illuminant A)
Temperature/humidity drift Within ±3% ±1 digit (of value displayed at 20°C/68°F) within operatingtemperature /humidity range
Calibration mode Minolta standard/user-selected standard (switchable)
Color correction factor Set by numerical input; range: 0.001 to 9.999
Reference luminance 1; set by measurement or numerical input
Measurement modes Luminance; luminance ratio; peak luminance or luminance ratio
Display External: 4-digit LCD with additional indications
Viewfinder: 4-digit LCD with LED backlight
Data communication RS-232C; baud rate: 4800bps
External control Measurement process can be started by external device connected to data output terminal
Power source One 9V battery
Power consumption

While measuring button is pressed and viewfinder display is lit: 16mA average

While power is on and viewfinder display is not lit: 6mA average

Operating temperature/humidity range 0 to 40°C, relative humidity 85% or less (at 35°C) with no condensation
Storage temperature /humidity range -20 to 55°C, relative humidity 85% or less (at 35°C) with no condensation
Dimensions 79×208×150mm (3-1/8×8-3/16×5-7/8 in.)
Weight 850g (30 oz.) without battery
Standard accessories Lens cap; Eyepiece cap; ND eyepiece filter; 9V battery; Case

Specifications are subject to change without notice.

* Equivalent to 2% specified for T-1 series.
8% CIE(f1'),new JIS(1993)
2% old JIS